[:SENSe]:CORRection:COLLect:INFO:STATus?
Query Description: Query returns 0 if calibration is not in progress, 1 if calibration is in
Example: To query calibration progress status:
[:SENSe]:CORRection:COLLect:INITialize
Cmd Description: Command to initialize an RF calibration with the current calibration
Example: To initialize RF calibration:
[:SENSe]:CORRection:COLLect:LOAD
Cmd Description: Perform 'Load' step of calibration.
Example: To measure load:
[:SENSe]:CORRection:COLLect:OPEN
Cmd Description: Perform 'Open' step of calibration.
Example: To measure open:
[:SENSe]:CORRection:COLLect:SAVe
Cmd Description: Complete an RF Cal and perform coefficient calculation.
Example: To complete a RF calibration and perform coefficient calculation:
[:SENSe]:CORRection:COLLect:SHORt
Cmd Description: Perform 'Short' step of calibration.
Example: To measure short:
4-24
Chapter 4 — Cable & Antenna Analyzer Mode Commands
progress, 2 if calibration was aborted, 3 if cal-coefficients are being
calculated, or 4 if calibration has completed.
:CORR:COLL:INFO:STAT?
type.
:CORR:COLL:INIT
:CORR:COLL:LOAD
:CORR:COLL:OPEN
:CORR:COLL:SAV
:CORR:COLL:SHOR
PN: 10580-00322 Rev. F
S331L PM
Need help?
Do you have a question about the Site Master S331L and is the answer not in the manual?