Analog Output; Setup; Measurement Conversion - MICRO-EPSILON optoCONTROL 2600 Operating Instructions Manual

Table of Contents

Advertisement

Measure
optoCONTROL 2600
6.4

Analog Output

6.4.1

Setup

Setup occurs specific to the measurement program in the menu Edit program:
> Enter gain for analog output or Enter gain for analog output.
6.4.2

Measurement Conversion

The measurement value (MV) is calculated from the analog output voltage as follows:
4.0
MV (mm)
Analog gain
Value ranges
Analog offset: -50.0000 V ... +50.0000 V
Analog gain:
-4.00000 ... +4.00000
The analog output voltage to be expected for a certain measurement value can be calcu-
lated from the following formula:
MV (mm)
V
(V)
OUT
4.0
With the two quantities, analog gain and analog offset, you can produce all the arising
linear output characteristics as shown in the following graph. This is particularly interest-
ing for adaptation to evaluation equipment with lower resolution or lower voltage span on
the input. In this respect, the above formula is changed according to the analog gain.
Then it is possible, for example, to extend a measurement span of 10 mm to a voltage
span of 10 V; the analog gain in this case is +4.0.
10
5
0
-5
-10
0
x in mm
Fig. 38 Analog scaling of output characteristics
i
Zero-setting, see 6.3.7.1
It should therefore be carried out before the analog gain is changed.
The output voltage has an overrun respectively underrun of 20 mV (0.068 mm). This
means it can exceed resp. undercut the zero point and the full scale (+10 VDC) by
20 mV in each case.
In the above example with AG 2 and AO 0 V the output voltage from an edge position x
> 20 mm is limited to 10.02 V. With x > 40,.... mm the error value of 10.04 V then ap-
pears.
If negative output voltages disturb, an analog offset of -20 mV (-0.020 V) can be entered.
Error
With
in the standard setting, a voltage of +10.04 V is output.
(V
- Analog offset)
OUT
*Analog gain + Analog offset
AF
20
40
AO
and
mastering, see 6.3.7.2
AF -1, AO 10 V
AF 1, AO 0 V
AF 2, AO 0 V
AF 1, AO -5 V
AF 2, AO -10 V
AF 4, AO -25 V
Analog gain
Analog offset
also affect the analog output.
Page 36

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the optoCONTROL 2600 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

This manual is also suitable for:

Odc 2600-40Odc 2600-40(209)

Table of Contents