MICRO-EPSILON optoCONTROL 2600 Operating Instructions Manual page 40

Table of Contents

Advertisement

Measure
Binary format
Data conversion
Start
0
0
6 Bit
(D5...D0)
Fig. 43 Transmission format of a data word (example)
L-Byte
0
0
D5
M-Byte 0
1
D11
H-Byte 1
0
My
Fig. 44 Reception
My
Mx
Seg. 1
0
0
Seg. 2
0
1
Seg. 3
1
0
Seg. 4
1
1
Fig. 45 Measurement allocation for Multi-segment operating mode
D15
D14
D13
D12
Fig. 46 Result of conversion (digital value DW)
The following formula is used for the conversion of the digital values (DV) to the measurement values (MV in mm):
MV (mm)
DV * 40.824 / 65519 - 0.4204872
The settings Offset and Gain for the display or the analog output have no effect on the digital value.
With errors a digital value of
Error messages during measurement output:
65521
No edge
65522
At the beginning of the picture
65523
At the end of the picture
65524
Dark - bright edge
65525
Bright - dark edge
65526
Min. number of edges
65527
Max. number of edges
65528
Invalid measuring program
65529
Segment 1st edge > 2nd edge
65530
Segment number of edges < last edge
65531
Invalid working distance
65533
Laser off
65534
Invalid float
65535
DMA setup error
optoCONTROL 2600
Stop Start
0
D4
D3
D2
D1
D10
D9
D8
D7
Mx
D15
D14
D13
D11
D10
D9
D8
65520 is output.
1
6 Bit
Stop Start
(D11...D6)
D0
D6
D12
D7
D6
D5
D4
1
0
6 Bit
(My...D12)
D3
D2
D1
D0
Stop
Page 40

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the optoCONTROL 2600 and is the answer not in the manual?

Questions and answers

This manual is also suitable for:

Odc 2600-40Odc 2600-40(209)

Table of Contents