6.6.3.12 Read Measurement Program Data - MICRO-EPSILON optoCONTROL 2600 Operating Instructions Manual

Table of Contents

Advertisement

Measure
optoCONTROL 2600

6.6.3.12 Read Measurement Program Data

Command:
Byte 1
Byte 2
„+"
„+"
„0"
„D"
0x26
0x20
Response:
Byte 1
Byte 2
„0"
„D"
0x26
0xA0
0x26
0xE0
Measurement program
number
Measurement program
name
Place-holder
Analog offset
Analog gain
Display offset
Display gain
Upper limit
Lower limit
Upper warning
Lower warning
Reserve 1
Measurement mode
Median
Number of averages
Reserve 2
Measurement program
Number of segments
Front edge for Segment
1+2
Front edge for Segment
3+4
Reserve 4
Reserve 5
Rear edge for Segment
1+2
Rear edge for Segment
3+4
Reserve 7
Reserve 8
Place-holder
Master value
Fig. 50 Measurement program data
Byte 3
Byte 4
„+"
0x0D
0x0D2B2B2B
„C"
„1"
0x00
0x00
Byte 3
Byte 4
„C"
„1"
0x16
0x00
0x03
0x00
Binary - 0x0007
ASCII - 0x45
0x44
0x47
0x45
0x48
0x4C
0x55
0x00
Binary - 0x0000...0xFFFF
Binary - 0x00000000
Binary - 0x3F800000
Binary - 0x00000000
Binary - 0x3F800000
Binary - 0x42200000
Binary - 0x00000000
Binary - 0x42200000
Binary - 0x00000000
Binary - 0x0000
Binary - 0x0000
Binary - 0x0003
Binary - 0x0001
Binary - 0x0000
Binary - 0x0001
Binary - 0x0001
Binary - 0x0000
Binary - 0x0000
Binary - 0x0000
Binary - 0x0000
Binary - 0x0000
Binary - 0x0000
Binary - 0x0000
Binary - 0x0000
Binary - 0x0000...0xFFFF
Binary - 0x00000000
hex
0x3143444F
0x00002026
hex
Remark
0x3143444F
0x0003A026
Without error
0x0003E026
Error + 4 bytes
error code
USER2-Measurement program
„E"
„D"
„G"
„E"
„H"
„L"
„U"
ZERO
0.0 VDC
1.0
0.0 mm
1.0
40.0 mm
0.0 mm
40.0 mm
0.0 mm
NORMAL
over 3 values
1
Edge HL
Edge HL, LH, Diameter, Gap
0,0 mm
Page 51

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the optoCONTROL 2600 and is the answer not in the manual?

Questions and answers

This manual is also suitable for:

Odc 2600-40Odc 2600-40(209)

Table of Contents