Measure
optoCONTROL 2600
6.6.3.3
Stop Command
STOP
Name
Description:
Stops the permanent measurement output from the sensor.
Command:
Byte 1
Byte 2
„+"
„+"
„0"
„D"
0x21
0x20
Response:
Byte 1
Byte 2
„0"
„D"
0x21
0xA0
0x00
0x00
i
Start is on when the sensor is switched on. The command Stop is volatile and is
lost when the voltage supply is switched off or the Reset command is sent.
6.6.3.4
Reset Command
RESET
Name
Description:
The sensor executes a software reset. This corresponds to switch-
ing the sensor off and then on again.
Command:
Byte 1
Byte 2
„+"
„+"
„0"
„D"
0x01
0x20
Response:
Byte 1
Byte 2
„0"
„D"
0x01
0xA0
0x00
0x00
Byte 3
Byte 4
„+"
0x0D
0x0D2B2B2B
„C"
„1"
0x00
0x00
Byte 3
Byte 4
„C"
„1"
0x03
0x00
0x00
0x00
Byte 3
Byte 4
„+"
0x0D
0x0D2B2B2B
„C"
„1"
0x00
0x00
Byte 3
Byte 4
„C"
„1"
0x02
0x00
0x00
0x00
hex
0x3143444F
0x00002021
hex
Remark
0x3143444F
0x0003A021
Without error
0x00000000
Error code
hex
0x3143444F
0x00002001
hex
Remark
0x3143444F
0x0002A001
Without error
0x00000000
Error code