Technical Specifications - KLS Martin group MCO25plus Operating Instructions Manual

Co2 laser system
Table of Contents

Advertisement

11

Technical Specifications

Type of laser:
Laser wavelength:
Laser output power:
Laser class:
Classification (device class): IIb acc. to Medical Devices Act / Medical Devices Directive (MDD)
Protection class:
Type of protection:
Operating modes:
Mode structure:
Laser beam divergence:
Laser tube:
Minimum tube life:
Pilot laser:
Pilot laser power:
Laser beam delivery:
Scope of delivery:
Standard applicator:
Alternative applicators:
Focus diameter:
Cooling:
Heat exchange:
Ambient temperature:
88
CO
laser
2
λ = 10,600 nm
2–25 W ± 20%
IV
I
IP X1
1)
Continuous (cw)
2)
Pulsed operation:
single pulse
pulse train
average output power:
superpulse
3)
Cyclical (cyclically repeated pulses)
Cycle length: continuous, 10 ms to 10 s
TEM 00
2.5 ± 0.5 mrad (without application handpiece)
sealed off, DC-excited
2,000 h (maximum 24 months)
635 nm, class 3B diode laser
2 mW, continuously adjustable
Articulated spring arm with 7 joints/mirrors,
arm length = 1,300 mm, handpiece exchangeable
Articulated mirror arm w. 7 mirrors, footswitch and standard
applicator
Focusing handpiece, f = 127 mm
Focusing handpieces, f = 50 mm and 200 mm;
Soft Scan Plus scanner; scanner handpieces: 127 mm and 200 mm
0.1 mm at a focal distance of 50 mm
0.20 mm at the standard focal distance of 127 mm
0.32 mm at a focal distance of 200 mm
closed water circuit
by air
10–45°C (50–113°F)
MCO25plus Operating Instructions
2 to 25 W
5 ms – 10 s;
output power: 10–25 W
Frequencies adjustable 0.1–100 Hz
minimum pulse length: 5 ms;
< 0.1–25 W
0.3 ms; 0.3–11 W
V 4.1