5.7

Display Content Explanation

No.
Designation
D1
Output power
(F)
"Max. power"
D2
Pulse length
(F)
"Pulse"
"Figure size"
D2
(pattern size)
(C)
D3
Pause duration
TM
"Pause"
D3
Pulse rate
FM
"Pulse rate"
D3
Average
(F4)
superpulse power
D3
"Scan speed"
(C)
(scanner)
D4
"Cycle length"
D4
"Handpiece"
(C)
D5
"Cycle pause"
TM
D5
"Cycle rate"
FM
D5
"Center"
(C)
D6
"Pilot"
D7
"Average power"
(F)
D7
"Energy density"
(C)
V 4.1
Function
Laser output power in watts;
indicates peak power in pulsed or
multi-pulse (pulse-train) mode
Duration of one pulse in ms or s
Side length or diameter of the
scanning pattern (in mm)
Interval between two pulses (=
interpulse period), indicated in s
Pulse frequency/rate; number of
pulses per second (in Hz)
Superpulse output power
Selects the speed of the laser
beam's scanning movement
Duration of one cycle (pulse
package/train) in ms or s
Selects a scanner handpiece
Period of time (interval) between
two cycles, indicated in ms or s
Cycle frequency/rate; number of
cycles per second (in Hz)
Shifts the center of the scanning
pattern
Pilot (aiming) laser beam intensity
Average output power value,
calculated from the following
parameters: pulse length, pause
and peak power
Laser energy per unit area; kept
constant when the scanning pattern
changes (either in type or size)
Explanation
2 W to 25 W cw (F1),
10 W to 25 W in pulsed
and scanner mode (F2, F3,
C1...C5)
5 ms to 10 s (F2, F3)
For sizes, see table
5 ms to 10 s (F3)
5 ms to 0.1 s (F4)
0.1 Hz to 5 Hz (F3)
0.1 Hz to 200 Hz (F4)
0.3 to 11 W
low, medium, high
0.1 s to 10 s (F3, F4)
Focusing handpieces
(127 mm and 200 mm),
micromanipulators and
special optics
0.1 s to 10 s (F3, F4)
10 ms to 10 s (C)
0.1 Hz to 5 Hz (F3, F4, C)
"OFF" or "ON" indicated
Brightness/intensity
adjustable between 1%
and 100%
0.1 W to 25 W;
automatically calculated
and displayed
Energy density per unit
area in J/cm
2
.
Automatically calculated
from scanning speed,
handpiece type and output
power