9.2.5

"Confirm"

Whenever the "laser ready" button is operated or parameters are changed, the laser unit
automatically performs an output power check. If the power exceeds the permissible tolerance
range of ±20%, the measured value is indicated in the D7 section of the display.
Fig. 9.1
Display information (example): measured value lower
D1
S3
The system prompts the user to confirm the measured output power by pressing
"LASER READY" (S3, Fig. 9.1) again if he/she wishes to proceed.
Upon confirmation, the "CURRENT AVERAGE POWER" value – which may deviate from the
selected power – is displayed.
Fig. 9.2
Current average power following confirmation
D1
The laser can now be used as usual with the current output power. Should the problem (or
request for confirmation of the current output power) occur repeatedly, please call the
Technical Service.
84
Selected power
"Laser ready"
Maximum power
D7
Measured power
D7
Current power
MCO25plus Operating Instructions
V 4.1