5
Measuring safety-related output parameters
[
]
5.1
Checking output power at 2, 5, 10, 15, 20 and 25 W, using an external power
meter with a measuring accuracy greater than ± 10%. Permissible tolerance:
±20 %.
Notice!
If safety-related defects are found during this safety check, the unit must be
repaired before it is used again!
Measures taken:
[
]
Technical Service contacted on: ..........................................................
[
]
Unit withdrawn from service on: .........................................................
[
]
Results entered in medical device logbook on: ...........................................
Tester:
Date:
............................................................
Signature: