scanning line can then be optimally adjusted either by fine-rotation or with the
micromanipulator's joystick.
Notice: No energy density per unit area is indicated for line scans. Since these
types of scanning pattern are used for surgical cutting, their energy density is
always high (as indicated by "HIGH" in the corresponding field).
6.6.6

Parameters in Scanner Mode

MAX. POWER
Adjustment range:
SCANNING PATTERN SIZE (FIGURE SIZE)
Use the "SELECT" button to highlight the "Figure Size" field, then increase or decrease the
value with the arrow buttons as appropriate.
Adjustment ranges:
Pattern
Triangle
Square
Rectangle
Hexagon
Circle
Ring, thick
Ring, thin
Line, straight
Line, curved
Table 6.1:
Scan pattern sizes available for the two scanner handpieces
Notice: The most important parameter for the laser's tissue impact is energy
density. Changing the size of the scanning pattern has no influence on the
selected energy density! In other words, the tissue impact of the laser beam
remains exactly the same when you select a different pattern size.
SCANNING SPEED
Selectable values:
All speeds are below the thermal relaxation time of the tissue (which is 0.7 ms). The
scanning speed doubles from "low" to "medium", and again from "medium" to "high". The
higher the scanning speed, the lower is the thermal impact on the tissue.
58
from 10 to 25 W in 1-W increments
Available pattern sizes depend on the scanner handpiece
connected to the scanner:
Scanner Handpiece
127 mm
2–8 mm
3–13 mm
2–8 mm
3–13 mm
4–8 mm
6–13 mm
2–8 mm
3–13 mm
2–8 mm
3–13 mm
6–8 mm
9–13 mm
4–8 mm
6–13 mm
1–6 mm
1–9 mm
1–6 mm
1–9 mm
low, medium, high
MCO25plus Operating Instructions
200 mm
V 4.1