4.5.1
Preparing the Laser Unit
To prepare the opening for installation of the articulated mirror arm, proceed as follows:
Remove (unscrew) the white protective cap, then remove the gray cover/cap next to the
opening for the articulated arm. Using a 5-mm hexagon socket wrench (2, Fig. 4.2a), rotate
the hexagon socket-head screw located below the opening counterclockwise. This moves the
clamping jaw (1, Figures 4.2a and 4.2b) aside (Fig. 4.2b) to allow installation of the articulated
mirror arm.
Fig. 4.2a
Opening the clamping mechanism
to allow articulated mirror arm installation
1
Clamping jaw
2
5-mm hex socket wrench
3
Thread for screwing the articulated mirror arm in place
4
Irrigation gas connector
5
Scanner cable connector socket
26
MCO25plus Operating Instructions
Fig. 4.2b Clamping mechanism open;
arm can now be installed
V 4.1