2.17

Nominal Ocular Hazard Distances (NOHD)

For the accessories permitted for use with this laser, the following safety clearances (NOHDs)
must be observed, as otherwise the limit values for permissible laser irradiation and irradiance
would be exceeded!
Accessory
Focusing handpiece, 50 mm
Focusing handpiece, 127 mm
Focusing handpiece, 200 mm
Special optics set for nasal
operations
Soft Scan Plus scanner
Table 2.1:
Safety distances (NOHD) for MCO25plus accessories
18
NOHD Specifications
3 m
Focal distance = 50 mm
5 m
Standard focusing
handpiece
7 m
Focal distance = 200
mm
14 m
Cannula, straight or
angled
5 m
Handpiece, f = 127 mm
7 m
Handpiece, f = 200 mm
MCO25plus Operating Instructions