Technical Description; Technical Data - KLS Martin group LIMAX 120 Operating Instructions Manual

Surgical laser unit
Hide thumbs Also See for LIMAX 120:
Table of Contents

Advertisement

Operating Instructions
Surgical Laser Unit Limax
9

Technical Description

9.1

Technical Data

Designation
Type of laser
Laser wavelength
Laser output power
Pulse types
Pilot laser wavelength
Pilot laser power
Laser beam delivery
Laser beam quality
Laser fiber connector
Control / monitoring
Operation
Cooling
Mains connection,
version E (U)
Mains current, version E (U)
Line fuses, version E (U)
Power input
Laser class
Class of protection
Type of protection
Classification acc. to MDD
Pilot laser
Sound level (operating noise)
Smoke evacuator (VAC)
VAC control
V. 2.0
®
120
Specification Limax
Diode-pumped Nd:YAG laser
1.320 nm
5 – 120 W
Continuous (cw)
Single pulse: pulse on-time:
Pulse train, adjustable: pulse on-time:
635 nm
5 mW, continuous-wave (CW) or pulsed
Brightness setting range: 2 % – 100 %
Laser fibers, laser focusing handpiece
Numerical laser aperture 0.22
socket (mechanically coded SMA socket)
SMAplus
2 microprocessors
Rotary switch and membrane buttons, 8.4'' color display
Compressor air cooling
230 – 240 VAC; 50 / 60 Hz (100 – 240 VAC; 50 / 60 Hz)
max. 13 A (max. 26 A)
2 x T 16 A H, 500 V and 2 x F 20 A (2 x T 16 A H, 500 V
and 2 x F 30 A)
max. 3000 VA
4
I
IP X1
II b
3R
Sound level: 51 dB(A), at full load: 60 dB(A)
integrated
MCB (Martin Communication Bus), control via Limax
®
120
pulse off-time:
0.1 s – 10 s
0.1 s – 10 s
0.1 s – 10 s
77

Advertisement

Table of Contents
loading

Table of Contents