MCO25plus Operating Instructions
4.4.5

OR Door Interlock

An external door contact may also be integrated into the safety system. When using this
device, the laser immediately goes into "STANDBY" mode as soon as the OR door is opened,
and the irradiation process is interrupted at once.
After closing the interlock contact again, the laser must first be switched to "ready" before the
treatment can be resumed. The door contact must be a floating contact (i.e. potential-free),
rated for a load of 24 VDC/0.1 A. It is connected to the laser unit via a 3-pole circular
connector that must be plugged into socket 5 (Fig. 4.1).
If you do not wish to use this safety device, the jumper plug supplied with the laser must be
inserted into socket 5 (Fig. 4.1). This plug has an internal contact bridge.
Warning: Never insert the shorting plug into the footswitch socket!
4.4.6

Equipotential Bonding Pin

This connector (6, Fig. 4.1) must be used if the operating room's electrical installation requires
separate grounding of the unit.
4.4.7

Fuses Located in Mains Filter

After opening the hinged cover 7 (Fig. 4.1), two miniature fuses are accessible whose ratings
are indicated on the holder.
4.5

Connecting the Articulated Mirror Arm

Attention!
The jointed mirror arm is a precision component. Therefore, be sure to keep it
free from mechanical loads when moving or transporting it.
V 4.1
25