MCO25plus Operating Instructions
5.6

Function Key Explanation

No.
Designation
CW
F1
Continuous-wave
laser beam
F2
Single pulse
F3
Pulse train
F4
Superpulse
F5
Cycle
Triangular
C1
scanning pattern
Square/rectan-
C2
gular scanning
patterns
Hexagonal/circu-
C3
lar scanning
patterns
Circular scanning
C4
pattern (ring)
V 4.1
Function
Selects continuous-wave (CW) mode
Selects single-pulse mode; the laser
emits a single laser light pulse
Selects multi-pulse mode; the laser
emits laser light pulses on a
repetitive basis
Selects superpulse mode; the laser
emits very short but high-power
pulses on a repetitive basis
In "cycle" mode, the laser generates
pulse packages on a repetitive basis
when combining it with F3 or F4; if
used in connection with a "C" button,
the selected scanning pattern is
automatically repeated
Selects the scanning function;
pressing this button repeatedly
rotates the triangle by 90 degrees
counterclockwise
Selects the scanning function;
pressing this button repeatedly
rotates first the square, then the
rectangle by 45 degrees
Selects the scanning function;
pressing this button twice rotates the
hexagon by 30 degrees; pressing it
three times selects the circle
Selects the scanning function;
pressing this button repeatedly
selects a thin or thick ring
Explanation
Function key, yellow light;
laser stops as soon as the
footswitch is released
Function key, yellow light;
laser stops after emission
of one pulse
Function key, yellow light;
laser stops as soon as
footswitch is released;
can be combined with the
F5 cycle function
Function key, yellow light;
laser stops as soon as
footswitch is released;
can be combined with the
F5 cycle function
Function key, yellow light
Scanning button, yellow
light; single scan. (With
F5, the scan is repeated
periodically; laser stops
as soon as the footswitch
is released.)
Scanning button, yellow
light; single scan. (With
F5, the scan is repeated
periodically; laser stops
as soon as the footswitch
is released.)
Scanning button, yellow
light; single scan. (With
F5, the scan is repeated
periodically; laser stops
as soon as the footswitch
is released.)
Scanning button, yellow
light; single scan. (With
F5, the scan is repeated
periodically; laser stops
as soon as the footswitch
is released.)
39