10.2. Temperature Monitoring Messages

Note – Factory default Temperature alarm and limit settings may be found in the Menu
Settings table in the PPM3 Monitor Settings chapter.
Message
TEMP < [lower limit]
TEMP > [upper limit]
TEMP unplugged
TEMP out of range
TEMP problem detected
April 2019
Temperature Monitoring
Parameter
Possible Cause
Value
The patient's
temperature has fallen
[number]
below the current lower
alarm limit.
The patient's
temperature has risen
[number]
above the current upper
alarm limit.
Temperature probe
[blank]
disconnected
The patient's
temperature has risen
above the maximum
value the monitor can
---
accurately detect.
There is a problem with
the connections or with
the hardware.
Monitor has detected a
---
hardware problem.
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
Suggested Action
Check the patient and provide any
necessary clinical care.
Change the alarm limit if it is no longer
clinically appropriate.
Check the patient and provide any
necessary clinical care.
Change the alarm limit if it is no longer
clinically appropriate.
Check to make sure the temperature
probe is connected to the temperature
cable.
Check to make sure the temperature
cable is connected to the PPM3.
Check the patient and provide any
necessary clinical care.
Check the temperature cable
connections.
Turn the monitor off, then on.
If message persists, contact Zoe
technical support.
Turn the monitor off, then on.
If message persists, contact Zoe
technical support.