Cleaning
warm air for 2 hours, then check to make sure all monitoring functions are
still working properly.
CAUTION – Take particular care when cleaning the NBP cuff, NBP hose,
and NBP connector on the PPM to prevent fluid from entering the
connectors. Fluid in the NBP airway may affect blood pressure
determination accuracy and damage the monitor.
CAUTION – Microstream® ETCO
sampling lines are designed for single
2
patient use, and are not to be reprocessed. Do not attempt to clean,
disinfect, sterilize or flush any part of the sampling line as this can cause
damage to the monitor.
CAUTION – Accessories that fall on the floor should be inspected for
contamination and proper functionality. If contamination is observed, then
this cleaning procedure should be followed.
CAUTION – The user has the responsibility to validate any deviations from
the recommended method of cleaning and disinfection.
21-3
April 2019
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e