PPM3 Monitor Settings
Parameter Settings
Setting name
P2m Alarms On
P2m Print on Alarm
P2d Lower Alarm Limit
P2d Upper Alarm Limit
P2d Alarms On
P2d Print on Alarm
P2 Label
(P2 waveform) Size
P2 Format
TEMP Lower Alarm Limit (Note:
Depends On User Setting For
TEMP Units)
TEMP Upper Alarm Limit (Note:
Depends On User Setting For
TEMP Units)
TEMP Alarms On
TEMP Print on Alarm
NBPs Lower Alarm Limit
NBPs Upper Alarm Limit
NBPs Alarms On
NBPs Print on Alarm
NBPm Lower Alarm Limit
NBPm Upper Alarm Limit
NBPm Alarms On
NBPm Print on Alarm
NBPd Lower Alarm Limit
NBPd Upper Alarm Limit
NBPd Alarms On
NBPd Print on Alarm
NBP Interval
NBP Initial Inflation Pressure
Trend Interval
19-4
Default value
Adult / Pediatric
No
---
No
---
55 mmHg
---
100 mmHg
---
No
---
No
---
P2
---
0 to 200
---
mmHg
SYS/DIA
---
36.0 C
---
97.0 F
39.0 C
---
102.0 F
No
---
No
---
90 mmHg
---
180 mmHg
145 mmHg
No
---
No
---
75 mmHg
---
110 mmHg
---
No
---
No
---
55 mmHg
---
100 mmHg
---
No
---
No
---
Off
---
160
120 mmHg
mmHg
15 min
---
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
Possible values
Yes, No
Yes, No
-10 to 209 mmHg, Off
-9 to 210 mmHg, Off
Yes, No
Yes, No
P2, ART, PA, CVP
-10 to 20 mmHg, -10 to 60 mmHg, 0 to 150 mmHg,
0 to 200 mmHg, 0 to 250 mmHg, 0 to 300 mmHg
SYS/DIA, MEAN
32.0 to 41.9 degrees C, Off
89.6 to 107.5 degrees F, Off
32.1 to 42.0 degrees C, Off
89.7 to 107.6 degrees F, Off
Yes, No
Yes, No
30 to 249 mmHg, Off
31 to 250 mmHg, Off
Yes, No
Yes, No
20 to 229 mmHg, Off
21 to 230 mmHg, Off
Yes, No
Yes, No
10 to 209 mmHg, Off
11 to 210 mmHg, Off
Yes, No
Yes, No
Off, 1, 2, 3, 5, 10, 15, 30, 60, 120 minutes
100 to 270 mmHg in steps of 10 mmHg
1 min, 5 min, 15 min, 1 h, 4 h
April 2019