22. Maintenance and Storage

The following table shows the recommended maintenance procedures for the
Nightingale PPM3 monitor and its accessories. The Nightingale PPM3 monitor
does not require periodic recalibration – with the exception of the Oridion
Microstream® CO
module (see the PPM3 Calibration Procedures after the
2
Functional Tests). However, it is a good idea to check that the monitor is in good
working order, as described in the following table. These functional tests should
be done every 12 months, and they can be performed by clinicians or qualified
service personnel. The following equipment is needed to perform these
procedures:
• ECG/Respiration Simulator (e.g., Fluke Biomedical MPS-450)
• SpO
Simulator (e.g., Fluke Biomedical Index-2)
2
• Blood Pressure Simulator (e.g., Fluke Biomedical BP-Pump2)
• Safety Analyzer (e.g., Fluke Biomedical 505 Series)
• Oridion Microstream® calibration kit from Air Liquide (Scott
Medical); Part Number T4653ORFCD – including a canister of 5%
CO
, 21% O
2
and a calibration Filterline.

22.1. PPM3 Functional Tests

PPM3 Function
Mechanical Integrity
Knob
Front Panel Keys
Power LED
Battery Charging LED
April 2019
Maintenance and Storage
, with the balance N
2
2
Procedure
Check for cracks, abrasive edges and other signs of damage.
Verify menu navigation functionality by turning and pressing.
PRINT key: verify that a Vital Signs Report is generated at the MPC's printer.
ALARM SILENCE key: verify that alarm tone can be silenced.
NBP INTERVAL key: verify NBP Interval menu is displayed.
NBP START / STOP key: verify NBP measurement starts and stops.
STANDBY key: verify that unit enters Standby mode or displays Standby
Menu
SETUP key: verify that the top level Setup menu is displayed.
TRENDS key: verify that the Trends menu is displayed
MAIN SCREEN key: verify that the Main Screen is displayed
Verify that the green power LED is illuminated on the front of the PPM3.
Verify that the green charging LED is illuminated on the back of the PPM3
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
along with a T-piece connection
22-1