Oridion CO
Message
IPI age range not set
CO
Unplugged
2
CO
Occluded Line
2
CO
Purging Line
2
CO
problem detected
2
7-8
Monitoring
2
Parameter
Possible Cause
Value
For pediatric patients,
the age range must be
---
set to correctly determine
IPI.
CO
sampling line is not
2
---
connected.
CO
sampling line cannot
2
be cleared due to
---
moisture or other
obstruction.
Microstream module is
---
trying to clear excess
humidity
CO
interface has
2
---
encountered a problem.
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
Suggested Action
Set the age range in the CO
2
screen.
Connect the CO
sampling line to the
2
PPM3.
Disable CO
monitoring in the PPM3
2
Parameters menu if these parameters
are no longer clinically required.
Replace the sampling line. If connected
to scavenging system, disconnect to see
if message disappears.
If the message disappears, resume
usage. If the purging operation is
unsuccessful or the Occluded Line
message appears, disconnect the line
and reattach. If the problem continues,
replace the sampling line.
Check the CO
subsystem including
2
sampling line and exhaust port. Verify
that the exhaust port is not blocked.
April 2019