12.2. Alarm Conditions

The tables in this section contain lists of all the conditions the PPM3 can detect
for each parameter, along with alarm characteristics of the condition. The first
row in each table contains the "normal condition" for the parameter, and the other
rows contain the "alarm conditions" for the parameter.
Columns in these tables have the following meaning:
• Condition
the name of the condition
• Display value
is true (applies only to physiological parameters)
• Alarm grade
• Message
the text of a message displayed in the message area when the
condition is true
• Annunciation type
Note – The delay between alarm annunciation on the PPM3 and remote annunciation on
the MPC central station should be no more than 1 sec.
General Monitor Conditions:
Condition
Monitor okay
Monitor problem detected
MPC Communication Link Conditions:
Condition
MPC link okay
MPC link lost
April 2019
the value displayed for the parameter when the condition
as defined earlier in this chapter
as defined earlier in this chapter
Display
Alarm
Value
Grade
N/A
None
N/A
Low
Display
Alarm
Value
Grade
N/A
None
N/A
Low
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
Managing Alarms
Message
None
Monitor need service
Message
None
MPC connection lost
Annunciation
Type
Persistent
One-time
Annunciation
Type
Persistent
Persistent
12-9