ZOE PPM3 User Manual page 122

Nightingale monitoring system
Table of Contents

Advertisement

Working With Menus
Parameter
HR/PR, RR,
ETCO
, RRc
2
NBPs/m/d,
P1s/m/d,
P2s/m/d
FICO
2
IPI
SpO
2
Temp
11-22
Lower limit adjustment
Changed to 80% of the
current parameter value (or
the nearest allowable value
given alarm limit rules)
N/R
Changed to 75% of the
current parameter value (or
the nearest allowable value
given the alarm limit settings
rules)
Changed to 95% of the
current parameter value (or
the nearest allowable value
given alarm limit rules)
Changed to 95% of the
current parameter value (or
the nearest allowable value
given alarm limit rules)
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
Upper limit adjustment
Changed to 125% of the
current parameter value (or
the nearest allowable value
given alarm limit rules)
Changed to 125% of the
current parameter value (or
the nearest allowable value
given alarm limit rules)
N/R
Set to 100%
Changed to 105% of the
current parameter value (or
the nearest allowable value
given alarm limit rules)
April 2019