Nightingale monitoring system
Table of Contents

Advertisement

Units
Parameters
CO
Measurement Range
2
CO
Measurement Accuracy
2
CO
Resolution
2
RRc (Resp. Rate) Measurement
Range
RRc Resolution
Barometric Pressure Range
Barometric Pressure
Compensation
Report Interval
Warm-up Time Required to
Meet Accuracy Specifications
Total System Response Time
Drift of Measurement Accuracy
Measurement Accuracy for Gas
Mixture
Measurement Accuracy in the
Presence of Interfering Gases
Invasive Blood Pressure
Transducer Type
Transducer Excitation Voltage
Frequency Response
Measurement Units
Parameters
Measurement Range
Measurement Accuracy
April 2019
mmHg
ETCO
, FICO
, RRc
2
2
ETCO
& FICO
: 0 to 150 mmHg
2
2
Dry single gases at 22 ± 5°C and 1013 ± 40 hPa
0 to 15 vol%: ±(0.2 vol% + 2% of reading)
15 to 25 vol%: unspecified
All conditions
±(0.3 kPa + 4% of reading)
1 mmHg
0 to 150 bpm. RRc is displayed after 3 breaths and the average
value is updated every breath.
1 bpm
525 to 1200 hPa
Automatic
Per breath
< 10 seconds (concentrations reported and full accuracy)
< 1 second
Complies with EN ISO 21647:2004 standard
Complies with EN ISO 21647:2004 standard
Complies with EN ISO 21647:2004 standard
Strain gauge
5.00 VDC ± 1 %
0-12 Hz
mmHg
Diastolic, Systolic, Mean for all except Mean-only for CVP
-50 to 300 mmHg
± 1 mmHg or ± 1 %, whichever greater
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
Technical Data
24-9