Recommended separation distances between portable and mobile RF communications equipment
The Nightingale PPM3 is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The user of the Nightingale PPM3 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the Nightingale PPM3 as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
Note 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
24-16
and the Nightingale PPM3
Separation distance according to frequency of transmitter
150 kHz to 80
MHz
d = 1.2
0.12
0.38
1.2
3.8
12
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
(m)
80 MHz to 800
MHz
d = 1.2
0.12
0.38
1.2
3.8
12
800 MHz to 2.5
GHz
d = 2.3
0.23
0.73
2.3
7.3
23