Invasive Blood Pressure; Figure 38. Setup Ibp Menu - ZOE PPM3 User Manual

Nightingale monitoring system
Table of Contents

Advertisement

Working With Menus
11.1.7.

Invasive Blood Pressure

Setting
Description
Lower Limit
P1 systolic/mean/diastolic alarm lower limits
Upper Limit
P1 systolic/mean/diastolic alarm upper limits
Auto-adjust lower limits to 80% of current values and upper limits to
Auto
125% of current values.
Alarms On
Enables (Yes) or disables (No) alarms
Print on Alarm
Print a strip chart summary upon alarm conditions
Label
Changes label from default to ART, PAP, or CVP.
Size
Changes display amplitude of P1 waveform
Format
Displays either Sys/Dia or Mean numeric value in larger font
Zero Set
Zeroes P1 when transducer vent has been opened
11-12

Figure 38. Setup IBP Menu

N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
April 2019