Electrode Placement For Paced Patients (Iec); Figure 18. Paced Iec Electrode Placement - ZOE PPM3 User Manual

Nightingale monitoring system
Table of Contents

Advertisement

ECG and Heart Rate Monitoring

3.7. Electrode Placement for Paced Patients (IEC)

For ECG cable lead sets with IEC (Europe) lead designations, the optimal
electrode placement for patients with pacemakers may be as illustrated in the
following diagram:
Note – If using a three-wire ECG cable lead set, you only need to apply the R, L, and F
electrodes, and only Lead II will be available.
R (Red)
N (Black)

Figure 18. Paced IEC Electrode Placement

1. Position the right arm (R) electrode on the right mid-clavicular line, 5
intercostal space.
2. Position the left arm (L) electrode on the left mid-clavicular line, directly
below the clavicle.
3. Position the neutral (N) electrode on the right mid-clavicular line, 6
th
7
intercostal space (5-lead cable only).
4. Position the left leg (F) electrode on the left mid-clavicular line, 6
intercostal space.
5. Position the chest (C) electrode on the 4
border (5-lead cable only).
3-8
C (White)
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
L (Yellow)
F (Green)
th
intercostal space, right sternal
th
th
and
th
th
and 5
April 2019