24. Technical Data
General
Dimensions
Weight
Finish
Power Requirements
Mains Frequency Range
Power Consumption
Standards Conformance
Patient Risk Current
(IEC 60601-1)
Type of Protection (Electrical)
Degree of Protection (Electrical)
Degree of Protection (Water)
Disinfecting Method
Degree of Safety (Flammable
Anesthetic Mixture)
Mode of Operation
April 2019
11.3"W x 7.2"H x 2.4"D (288 mm x 182 mm x 60 mm)
4.5 lb (2.0 g), 5.0 lb w/ Oridion Microstream® CO
PC/ABS
100 – 240 VAC, 1.2 A max
50 – 60 Hz
12W nominal, 30W (when charging battery)
IEC 60601-1:2005 (General Safety)
AAMI ES60601-1:2005 (General Safety)
CSA C22.2#60601-1:2008 (General Safety)
IEC 60601-1-2:2007 (Class B) (EMC)
IEC 60601-1-4:2000 (General Safety)
IEC 60601-1-8:2006 (Alarms)
IEC 60601-2-27:2011 (ECG Monitoring)
AAMI SP10:2002/A1:2003 (Non-Invasive Blood Pressure)
ISO 80601-2-30:2009 (Non-Invasive Blood Pressure)
IEC 60601-2-34:2011 (Invasive Blood Pressure)
ISO 80601-2-55:2011 (CO
ISO 80601-2-56:2009 (Temperature)
ISO 80601-2-61:2011 (SpO
IEC 62366:2007 (Usability)
IEC 62304:2006 (Software)
Electromedical Apparatus with Isolated Patient Connection.
Meets the following limits:
Enclosure Risk Current < 100 μA
Patient-applied Risk Current < 10 μA
Patient Isolation Risk Current < 50 μA
Earth Risk Current < 500 μA
Class I
Type CF, Defibrillation-proof
Ordinary Equipment (IPX1)
Per the instructions in the Cleaning chapter
Not suitable for use in the presence of a Flammable Anesthetic
Mixture
Continuous
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
Technical Data
Respiratory Gas Monitoring)
2
)
2
(2.3 kg)
2
24-1