PPM3 Monitor Settings
Parameter Settings
Setting name
ETCO
Enabled
2
IBP Channels
Numbers Only
HR Lower Alarm Limit
HR Upper Alarm Limit
HR Alarms On
HR Print on Alarm
ECG Lead
(ECG Lead) Size
HR Source
Pulse Tone Source
ECG Filter
RR Lower Alarm Limit
RR Upper Alarm Limit
RR Alarms On
RR Print on Alarm
ETCO
Lower Alarm Limit
2
ETCO
Upper Alarm Limit
2
ETCO
Alarms On
2
ETCO
Print on Alarm
2
FICO
Upper Alarm Limit
2
FICO
Alarms On
2
FICO
Print on Alarm
2
RRc Lower Alarm Limit
RRc Upper Alarm Limit
RRc Alarms On
RRc Print on Alarm
(CO2 waveform) Size
Apnea Time
O
Compensation (Masimo)
2
19-2
Default value
Adult / Pediatric
No
---
None
---
No
---
50 bpm
---
120 bpm
150 bpm
Yes
---
No
---
II
---
10 mm/mV
---
ECG
---
Off
---
Monitor
(0.67 – 40
---
Hz)
5 bpm
10 bpm
20 bpm
45 bpm
Yes
---
No
---
35 mmHg
---
45 mmHg
---
Yes
---
No
---
2 mmHg
---
Yes
---
No
---
5 bpm
10 bpm
20 bpm
45 bpm
Yes
---
No
---
0 to 40
---
mmHg
20 seconds
---
0 to 30
---
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
Possible values
Yes, No
None, 1, 2
Yes, No
15 to 299 bpm, Off
16 to 300 bpm, Off
Yes, No
Yes, No
I, II, III, V, aVL, aVR, aVF
2.5 mm/mV, 5 mm/mV, 10 mm/mV, 15 mm/mV
ECG, SpO2, ART
ECG, SpO2, Off
Monitor (0.67 – 40 Hz),
Diagnostic (0.05 – 40 Hz)
5 to 149 bpm, Off
6 to 150 bpm, Off
Yes, No
Yes, No
0 to 145 mmHg, Off
5 to 150 mmHg, Off
Yes, No
Yes, No
0 to 98 mmHg, Off
Yes, No
Yes, No
5 to 149 bpm, Off
6 to 150 bpm, Off
Yes, No
Yes, No
0 to 20 mmHg, 0 to 40 mmHg, 0 to 60 mmHg, 0 to 80
mmHg
10, 15, 20, 25, 30 seconds
0 to 30, 31 to 70, 71 to 100
April 2019