ZOE PPM3 User Manual page 206

Nightingale monitoring system
Table of Contents

Advertisement

Technical Data
Measurement Accuracy for Gas
Mixture
Measurement Accuracy in the
Presence of Interfering Gases
TM
Masimo ISA
CO
Method
Units
Parameters
CO
Measurement Range
2
CO
Measurement Accuracy
2
CO
Resolution
2
RRc (Resp. Rate) Measurement
Range
RRc Resolution
Barometric Pressure Range
Barometric Pressure
Compensation
Report Interval
Flow Rate
Warm-up Time Required to
Meet Accuracy Specifications
Total System Response Time
Drift of Measurement Accuracy
Measurement Accuracy for Gas
Mixture
Measurement Accuracy in the
Presence of Interfering Gases
TM
Masimo IRMA
Method
24-8
Meets ISO 21647 Clause 51.101.3 (Tables 101 and 103):
± (volume fraction of 0.43% + 8% of gas level)
Meets ISO 21647 Clause 101.1 (Tables 101 and 105):
± (volume fraction of 0.43% + 8% of gas level)
/ Capnography
2
Sidestream (Non-dispersive IR)
mmHg
ETCO
, FICO
2
ETCO
& FICO
2
As measured with dry single gases:
0 to 15 vol%: ±(0.2 vol% + 2% of reading)
15 to 25 vol%: unspecified
1 mmHg
0 to 150 ± 1 bpm
1 bpm
525 to 1200 hPa
Automatic
Per breath
50 ± 10 sml/min
< 10 seconds (concentrations reported and full accuracy)
< 3 seconds (with 2m Nomoline sampling line)
Complies with EN ISO 21647:2004 standard
Complies with EN ISO 21647:2004 standard
Complies with EN ISO 21647:2004 standard
CO
/ Capnography
2
Mainstream
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
, RRc
2
: 0 to 150 mmHg
2
April 2019