Oridion Co Monitoring Messages - ZOE PPM3 User Manual

Nightingale monitoring system
Table of Contents

Advertisement

7.5. Oridion CO
Note – Factory default CO
Settings chapter.
Message
ETCO
< [lower limit]
2
ETCO
> [upper limit]
2
FICO
> [upper limit]
2
RR < [lower limit]
RR > [upper limit]
RR out of range (too high)
IPI < [lower limit]
CO
Warming Up
2
April 2019
Oridion CO
Monitoring Messages
2
alarm and limit settings may be found in the PPM3 Monitor
2
Parameter
Possible Cause
Value
The patient's ETCO
parameter value has
[number]
fallen below the current
lower alarm limit.
The patient's ETCO
parameter value has
[number]
risen above the current
upper alarm limit.
The patient's FICO
parameter value has
[number]
risen above the current
upper alarm limit.
[number]: The patient's
respiration rate (RRc)
[number]
has fallen below the
current lower alarm limit.
-or-
[0]
[0]: No breath has been
detected for the user-
configurable apnea time.
The patient's respiration
rate (RRc) has risen
[number]
above the current upper
alarm limit.
The patient's respiration
rate (RRc) has risen
above the maximum
value the monitor can
---
accurately detect.
Monitor confused by
signal artifact.
The patient's IPI
parameter value has
[number]
fallen below the current
lower alarm limit.
CO
module is preparing
2
---
to acquire data.
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
Monitoring
2
Suggested Action
Check the patient and provide any
2
necessary clinical care.
Change the alarm limit if it is no longer
clinically appropriate.
Check the patient and provide any
2
necessary clinical care.
Change the alarm limit if it is no longer
clinically appropriate.
Check the patient and provide any
2
necessary clinical care.
Change the alarm limit if it is no longer
clinically appropriate.
Check the patient and provide any
necessary clinical care.
Change the alarm limit if it is no longer
clinically appropriate.
Check the patient and provide any
necessary clinical care.
Change the alarm limit if it is no longer
clinically appropriate.
Check the patient and provide any
necessary clinical care.
Check the patient and provide any
necessary clinical care.
Change the alarm limit if it is no longer
clinically appropriate.
Allow more time.
7-7