Nightingale monitoring system
Table of Contents

Advertisement

Masimo CO
Monitoring
2
WARNINGS (IRMA) – Do not use the IRMA infant airway adapter with
adults as this may cause excessive flow resistance.
WARNINGS (IRMA) – Measurements can be affected by mobile and
RF communications equipment. It should be assumed that the IRMA
probe is used in the electromagnetic environment specified in this
manual.
WARNINGS (IRMA) – Do not place the IRMA airway adapter between
the endotracheal tube and an elbow as this may allow patient
secretions to block the adapter windows and result in incorrect
operation.
WARNINGS (IRMA) – To keep secretions and moisture from pooling
on the windows or oxygen sensor port, always position the IRMA
probe in a vertical position with the LED pointing upwards.
WARNINGS (IRMA) – Do not use the IRMA airway adapter with
metered dose inhalers or nebulized medications as this may affect
the light transmission of the airway adapter windows.
WARNINGS (IRMA) – Incorrect probe zeroing will result in false gas
readings.
WARNINGS (IRMA) – Incorrect agent selection by the user for IRMA
OR (no automatic agent identification) will result in false agent
readings.
WARNINGS (IRMA) – Replace the airway adapter if rainout or
condensation occurs inside the airway adapter.
WARNINGS (IRMA) – Use only PHASEIN manufactured IRMA airway
adapters.
WARNINGS (IRMA) – The IRMA probe is not intended to be in patient
contact.
CAUTION (ISA) – The ISA analyzer should be securely mounted in order
to avoid the risk of damage to the ISA.
8-9
April 2019
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e