Page 1
™ SELECT Multi-Parameter Monitor User Manual Addendum - ECG & Resp Parameters This User Manual Addendum describes the features and operations of Multi-Parameter monitor: Software Version 2.3 or above. SELECT 21-22-0335 Rev D Page 1 of 40...
Zoe Medical, Inc. ™ SELECT is a trademark of Zoe Medical, Inc. Refer to the SELECT User Manual, Zoe Medical PN 21-22-0316, for a list of trademarks. 21-22-0335 Rev D Page 2 of 40...
Overview SELECT TABLE OF CONTENTS Overview ......................2 TRADEMARKS ......................2 CONTACT ADDRESSES ..................... 3 CONVENTIONS USED IN THIS MANUAL ..............4 REVISION HISTORY ....................4 WARRANTY ........................ 4 INTENDED USE ....................8 INTENDED USE ......................8 OVERVIEW OF HEART RATE MONITORING ............8 OVERVIEW OF RESPIRATION RATE MONITORING ..........
INTENDED USE Full Indications for Use and Contraindications may be found in the User SELECT Manual, Zoe Medical PN 21-22-0316. monitor facilitates the monitoring of: SELECT ECG and heart rate (HR) with the ECG option for adult and pediatric patients;...
RR from the CO waveform. In this case the Resp/RR feature from ECG is disabled. Refer to User Manual Addendum - ETCO Parameter, Zoe Medical PN 21-22-0333 for more information. PEDIATRIC CONSIDERATIONS Cardiogenic artifact (heart rate pulsations that appear as "breaths" in the respiration signal) can be quite pronounced in Pediatric patients.
Additionally, proper placement of defibrillator paddles in relation to the ECG electrodes is required to minimize harm to the patient. Only use accessories approved by Zoe Medical for use with the monitor.
Read safety instructions provided with a defibrillator. The monitor is designed to withstand defibrillation and will recover within 5 seconds, per IEC 60601-1. Only use accessories approved by Zoe Medical for use with the monitor. Warning: Operating the monitor with HR/PR limit alarms disabled means that no low or high HR/PR alarm conditions will produce alarm notifications.
ECG MONITORING SELECT GETTING STARTED WITH ECG To begin monitoring ECG, use the following procedure: 1. Connect the ECG Patient Cable to the Monitor Locate the ECG Patient Cable Receptacle on the side of the Monitor. Refer to Figure 1. ECG Patient Receptacle on Monitor Receptacle key Figure 1: ECG Patient Cable Receptacle on left side of Monitor...
ECG MONITORING SELECT Insert the ECG Patient Cable to the ECG Patient Cable Receptacle located on the on the left side of the Monitor. Refer to Figure 2. Note: The ECG Receptacle is keyed so that the ECG Patient Cable can only be inserted into the monitor in one orientation.
ECG MONITORING SELECT 2. Connect the ECG Lead Set to ECG Patient Cable Connect an appropriate ECG 3-Lead Set (1 meter length) to the ECG Patient Cable (2 meter length). ECG 3-Lead Sets are available in AAMI or IEC colors, snaps (adult/pediatric), or mini-clip (neo) lead sets that to attach to the ECG electrodes.
ECG MONITORING SELECT CHECKING THE ECG SIGNAL When you have connected the patient following the steps listed above, you should be able to see a clean ECG signal similar to the Figure 4 below on the monitor display. If the ECG contains artifact or noise, review the steps for proper electrode site preparation and placement (see following pages).
ECG MONITORING SELECT ELECTRODE PLACEMENT STANDARD PLACEMENT (AHA) For ECG cable lead sets with AHA (USA) lead designations, position the electrodes as shown in Figure 5 below: RA (White) LA (Black) LL (Red) Figure 5: Standard AHA Electrode Placement 1. Position the right arm (RA) electrode on the right mid-clavicular line, directly below the clavicle.
ECG MONITORING SELECT PLACEMENT FOR PACED PATIENTS (AHA) For ECG cable lead sets with AHA (USA) lead designations, the optimal electrode placement for patients with pacemakers may be as illustrated in Figure 7 below: LA (Black) RA (White) LL (Red) Figure 7: Paced AHA Electrode Placement 1.
ECG MONITORING SELECT VERIFYING PROPER PACEMAKER HANDLING For paced patients, proper electrode placement results in pacer tick marks along the top of each ECG waveform channel. These tick marks occur at the point where the monitor detects a pacer pulse. The monitor rejects pacer pulses rather than calling them a QRS complex (per IEC 60601-2-27).
ECG MONITORING SELECT ECG SETUP To begin ECG monitoring use the following procedure: ECG LEAD WAVEFORM a. Touch the left side of the waveform area b. Touch the adjacent to the Waveform 1 selection (Refer to Figure 10); b. Touch I (II or III) and touch OK; and c.
ECG MONITORING SELECT ECG CASCADE Touch the left side of the waveform area Touch the Yes adjacent to the ECG Cascade selection to Enable; k. Touch the No adjacent to the ECG Cascade selection to Disable; Touch Home to return to the Main screen. Figure 11: ECG Cascade Enabled 21-22-0335 Rev D Page 20 of 40...
ECG MONITORING SELECT HR PARAMETER SETUP Table 11 on page 38 lists the Default HR Alarm Limits for Adult, Pediatric, and Neonatal. Table 12 on page 39 lists the Default ECG & HR Parameter Setting for Adult, Pediatric, and Neonatal. HR Alarm Limits and Parameter settings will operate for the current monitor patient mode.
ECG MONITORING SELECT Note: When Auto Limits adjust the Lower Limit, the adjustment shall not be allowed to be set to anything less than the Lower Limit, and when Auto Limits adjust the Upper Limit, the adjustment shall not be allowed to be set to anything greater than the Upper Limit. HR SOURCE The user-selectable HR Source feature allows the clinician to select the desired source of the Heart Rate numeric and associated alarm limit annunciation.
ECG MONITORING SELECT ANALYZE PACERS Touch the On adjacent to the Analyze Pacers selection to Enable. Touch the Off adjacent to the Analyze Pacers selection to Disable. When Analyze Pacers is set to On, the monitor will display white pace detect marks at the top of the ECG waveform to indicate where pacer pulses were detected.
ECG MONITORING SELECT ECG FILTER The user-selectable ECG Filter feature allows the clinician to select the desired filtering of the displayed ECG waveform. User Selectable settings: Monitor (0.67 - 40 Hz) (Default) & Diagnostic (0.05 - 40 Hz) CAL PULSE The user-selectable Cal Pulse feature allows the clinician to insert a 1 mV pulse onto the currently selected ECG waveform.
Inoperable ECG circuit wires. Intentional removal by Turn monitor off, then back on clinician If message persists, contact Zoe technical support. Press ALARM SILENCE in the event of intentional removal by clinician. Calm the patient. Isolate the patient from auxiliary equipment, if Patient movement possible.
The following Warnings and Cautions are directed toward the Respiration monitoring function. Additional Warnings and Cautions for the monitor are found in the SELECT User Manual, Zoe Medical PN 21-22-0316. SELECT WARNINGS Warning: Operating the monitor with RR limit alarms disabled means that no low or high RR alarm conditions will produce alarm notifications.
ACCESSORIES SELECT ENABLING DISPLAY OF RESPIRATION To enable the display of Respiration (Waveform and Numeric) use the following procedure: a. Touch the Setup button on the right side of the Main Screen b. Touch the Home Screen button and the Setup Home Screen will be displayed. Refer to Figure 15.
ACCESSORIES SELECT The Respiration Rate Monitoring settings and specifications may be found in the SPECIFICATIONS chapter of this manual. Procedures for changing configuration settings, such as enabling alarms, or adjusting alarm limits, may be found in the RESPIRATION SETUP chapter. FREEZING WAVEFORMS In some clinical environments the user may want to freeze the displayed waveforms.
ACCESSORIES SELECT RR PARAMETER SETUP Table 11 on page 38 lists the Default RR Alarm Limits for Adult and Pediatric. Table 12 on page 39 lists the Default RR Parameter Setting for Adult and Pediatric. RR Alarm Limits and Parameter settings will operate for the current monitor patient mode. RR ALARM LIMITS: 1) Touch RR Numeric field and the Setup RR menu will be displayed (refer to Figure 18);...
ACCESSORIES SELECT Note: When Auto Limits adjust the Lower Limit, the adjustment shall not be allowed to be set to anything less than the Lower Limit, and when Auto Limits adjust the Upper Limit, the adjustment shall not be allowed to be set to anything greater than the Upper Limit. TESTING RR ALARMS To manually test Respiration monitoring alarm functionality on a daily basis, you may use the following approach:...
Press ALARM SILENCE in the event of intentional Intentional removal by removal by clinician. clinician Turn monitor off, then back on If message persists, contact Zoe technical support. Calm the patient. Patient movement Isolate the patient from auxiliary equipment, if Electrical noise from possible.
Refer to Warnings and Cautions at the start of the User Manual, Zoe Medical PN 21-22-0316 for further precautions. SELECT GUIDANCE Contact your local waste disposal agency for guidance on the proper recycling or disposal of these components.
MAINTENANCE SELECT 7. MAINTENANCE The following table shows the recommended maintenance procedures for the SELECT monitor and its accessories. The monitor does not require periodic recalibration. However, it is a good idea to check that the monitor is in good working order, as described in the following table. These functional tests should be done every 12 months, and they can be performed by clinicians or qualified service personnel.
CLEANING SELECT 8. CLEANING Caution: Always disconnect the monitor from AC mains before cleaning. Do not use harsh chemicals for cleaning – in particular, do not use disinfectants that Caution: contain phenol as they can spot plastics. Do not steam autoclave, gas sterilize, irradiate, subject to intense vacuum, or immerse in water or cleaning solution.
SPECIFICATIONS SELECT 9. SPECIFICATIONS ECG AND HR Table 9 indicates the Specifications associated with ECG and HR Parameter. DEVICE MARKINGS Type CF Equipment (Defibrillation-proof) 3 wire ECG connector GENERAL Patient Populations Adult and Pediatric Accessories 3-lead cable Input Connector 7-pin connector Displayable Leads 3-lead cable: I, II, III HR Resolution...
SPECIFICATIONS SELECT RESPIRATION AND RR Table 10 indicates the Specifications associated with Respiration and RR Parameter. Patient Populations Adult, adolescent, child and infants Method Impedance Pneumography Input Connector Same as ECG Sensing Lead RR Resolution 1 rpm (breaths per minute) Measurement Range 0 to 120 rpm Measurement Accuracy...
SPECIFICATIONS SELECT Table 12 lists the ECG, HR & RR Parameter settings and Ranges. Defaults Parameter Range Adult Pediatric Neonatal ECG Lead I, II, III (ECG Lead) Size 10 mm/mV 10 mm/mV 10 mm/mV 2.5, 5, 10, 15 mm/mV HR Source Auto Auto Auto...
Page 40
SPECIFICATIONS SELECT NOTES: 21-22-0335 Rev D Page 40 of 40...