Technical Data
Sweep Speed
Display Mode
ECG
Accessories
Input Connector
Displayable Leads
HR Resolution
Measurement Range
Measurement Accuracy
Response Time
Report Interval
HR Averaging Scheme
Time To Alarm - Tachycardia
Notch Filter Frequency
Filter Bandwidth
Dynamic Range AC
Dynamic Range DC
Electrode Impedance
Defibrillation Protection
Pacer Pulse Detection
Pacer Pulse Rejection
Tall T-Wave Rejection
HR Response to Irregular
Rhythm
24-4
6.25, 12.5, 25 mm/s
Eraser Bar
3-lead cable, 5-lead cable
7-pin connector
3-lead cable: I, II, III, AVL, AVR, AVF
5-lead cable: I, II, III, AVL, AVR, AVF, V
1 bpm (beats per minute)
15 to 300 bpm
±2 bpm or ±1%, whichever is greater
Per IEC 60601-2-27, change from 80 to 120 bpm: < 7 seconds
Per IEC 60601-2-27, change from 80 to 40 bpm: < 11 seconds
1 second
Average of the 10 most recent, valid R-R intervals, discarding
the shortest and longest interval
IEC 60601-2-27, ECG Complex B1: < 10 sec (5 sec typical)
IEC 60601-2-27, ECG Complex B2: < 10 sec (9 sec typical)
50Hz, 60 Hz, Off
Monitor Mode: 0.67 Hz to 40 Hz (-3 dB)
Diagnostic Mode: 0.05 to 40 Hz (-3 dB)
±5 mV, per IEC 60601-2-27
±300 mV, per IEC 60601-2-27
>2.5 MOhm , per IEC 60601-2-27
Complies with IEC 60601-2-27
Lead II, I and V
Rejects all pulses of amplitude ±2mV to ±700mV and duration
0.1 to 2 ms, per IEC 60601-2-27, Clause 201.12.1.101.13
Rejects T-Waves less than or equal to 120% of a 1mV QRS and
a Q-T interval of 350 ms, per IEC 60601-2-27, Clause
201.12.1.101.17
IEC 60601-2-27, ECG Complex A1: HR is 80 bpm
IEC 60601-2-27, ECG Complex A2: HR is 65 bpm
IEC 60601-2-27, ECG Complex A3: HR is 120 bpm
IEC 60601-2-27, ECG Complex A4: HR is 91 bpm
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
April 2019