ZOE PPM3 User Manual page 149

Nightingale monitoring system
Table of Contents

Advertisement

SpO
Conditions:
2
Condition
SpO
within limits
2
SpO
< LL
2
SpO
> UL
2
SpO
Bad Probe
2
SpO
Cannot regulate LED
2
intensity (after start-up/standby)
SpO
Cannot regulate LED
2
intensity (after finger in probe)
SpO
Pulsations Too Weak
2
SpO
Probe is disconnected
2
(after start-up/standby)
SpO
Probe is disconnected
2
(after finger in probe)
SpO
motion artifact
2
PR Conditions:
Note ‒ When PR is sourced from SpO
in the HR parameter box. When PR is sourced from an ART-labeled IBP, the PR is
labeled in red as "HR (ART)." PR alarm conditions annunciate as HR alarms.
Condition
PR within limits
PR < LL
PR > UL
PR Bad Probe
PR Cannot regulate LED
intensity (after start-up/standby)
PR Cannot regulate LED
intensity (after finger in probe)
PR Pulsations Too Weak
PR Probe is disconnected
(after start-up/standby)
PR Probe is disconnected
(after finger in probe)
PR motion artifact
April 2019
Managing Alarms
Display
Alarm
Value
Grade
<NUM>
None
<NUM>
Medium
<NUM>
Medium
---
Low
<Blank>
None
---
Low
---
Low
<Blank>
None
---
Low
---
Low
, the Pulse Rate is labeled in cyan as "HR (SpO2)"
2
Display
Alarm
Value
Grade
<NUM>
None
<NUM>
Medium
<NUM>
Medium
---
Low
<Blank>
None
---
Low
---
Low
<Blank>
None
---
Low
---
Low
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
Message
None
SpO
< LL
2
SpO
> UL
2
SpO
replace sensor
2
None
SpO
check sensor
2
placement
SpO
weak signal
2
None
SpO
unplugged
2
SpO
artifact
2
Message
None
HR < LL
HR > UL
SpO
replace sensor
2
None
SpO
/ HR check sensor
2
placement
SpO
/ HR weak signal
2
None
SpO
/ HR unplugged
2
SpO
artifact
2
Annunciation
Type
None
Persistent
Persistent
One Time
Persistent
Persistent
Persistent
Persistent
One Time
Persistent
Annunciation
Type
None
Persistent
Persistent
One Time
Persistent
Persistent
Persistent
Persistent
One Time
Persistent
12-11