If you select the "Alarm Pause" item, the PPM3 will temporarily disable all alarm
monitoring. Selecting "Alarm Resume" from the same spot on the Setup Menu
will resume alarm monitoring. An "Alarms Paused" condition is indicated in the
message area with a timer indicating when alarm monitoring will resume. There
is also an alarms paused symbol displayed to the left of the message area.
Figure 41. Main Screen With Alarms Paused Message & Timer
WARNING – Pausing all alarms pauses all alarm conditions including
lethal arrhythmias (Asystole and VFib). Use this feature with extreme
caution as alarm checking is suspended for all alarm conditions.
Patients must be closely observed if all alarms are suspended.