ZOE PPM3 User Manual page 187

Nightingale monitoring system
Table of Contents

Advertisement

PPM3 Function
Oridion Microstream® CO
2
Masimo IRMA and ISA CO
2
Masimo ISA (Sidestream) CO
April 2019
Maintenance and Storage
Procedure
NOTE: Use only certified calibration gas apparatus that has not reached its
expiration date. This cal gas can be applied to the PPM3 patient monitor in
pulses that simulate patient breaths.
1. Connect the supplied CO
cal gas canister.
2. Pulse the calibration gas actuator, holding it depressed for 10 seconds and
released for 10 seconds (3 BPM). This will allow enough time for the
ETCO
to stabilize in both inhaled and exhaled states.
2
3. Verify that the ETCO
value reads 38 ± 2 mmHg and the FICO
2
reads 0 ± 2 mmHg.
NOTE: Factor a ±0.05 correction for every 100mmHg of ambient barometric
pressure above or below sea level respectively.
NOTE: The PPM3 should be returned to Zoe Medical for periodic servicing of
the CO
system after 30,000 hours of CO
2
NOTE: Use only certified calibration gas apparatus that has not reached its
expiration date. This cal gas can be applied to the PPM3 patient monitor in
pulses that simulate patient breaths.
1. Connect the PPM3 to a calibration gas canister (5% CO
Balance N
)
2
2. Pulse the calibration gas actuator, holding it depressed for 10 seconds and
released for 10 seconds (3 BPM). This will allow enough time for the CO
values to stabilize.
3. Verify that the FiCO
value is 0-2 mmHg and that the ETCO
2
within the range that is defined in the following table:
Current Atmospheric Pressure
mmHg
600 - 619
620 - 639
640 - 659
660 - 679
680 - 699
700 - 719
720 - 739
740 - 759
760 - 779
780 - 799
1.
Connect a new Nomoline sampling line with male luer lock to the ISA
2
LEGI and check that the LEGI shows a steady green light.
Connect a short silicon tubing with an inner diameter of 3/32" (2.4 mm)
2.
to the Nomoline male luer.
3.
Exhale a long breath into the silicon tubing until the CO2 concentration
is greater than 34 mmHg.
4.
Quickly connect the silicon tubing tightly to the ISA exhaust port.
5.
Wait 1 minute until the CO2 concentration has stabilized. Note the
value.
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
calibration FilterLine between the PPM3 and the
2
use.
2
Expected etCO2
mbar
mmHg
800 - 825
28 - 33
826 - 852
29 - 34
853 - 879
30 - 35
880 - 905
31 - 36
906 - 932
32 - 37
933 - 959
33 - 38
960 - 985
34 - 39
986 - 1012
35 - 40
1013 - 1039
36 - 41
1040 - 1065
37 - 42
value
2
, 21% 0
,
2
2
2
value is
2
22-3