Ecg And Heart Rate Monitoring Messages; Table 2: Hr Monitoring Messages - ZOE 740 SELECT User Manual

Multi-parameter monitor
Table of Contents

Advertisement

ECG MONITORING

ECG AND HEART RATE MONITORING MESSAGES

Table 2 indicates the Monitor Messages associated with the HR parameter:
Message
HR < [lower limit]
HR > [upper limit]
HR asystole
HR ventricular
fibrillation
HR lead off
HR artifact
21-22-0335 Rev D
Parameter
Possible Cause
Value
The patient's heart rate
has fallen below the
[number]
current lower alarm
limit.
The patient's heart rate
has risen above the
[number]
current upper alarm
limit.
No QRS detected for last
ASY
4 seconds
No organized ventricular
VF
rhythm detected
Unplugged cable
Broken cable
Loose lead wire
Faulty lead wire
[blank]
Dried out electrode
Inoperable ECG circuit
Intentional removal by
clinician
Patient movement
Electrical noise from
---
auxiliary equipment
Bad electrode contact

Table 2: HR Monitoring Messages

Suggested Action
Check the patient and provide any necessary
clinical care.
Change the alarm limit if it is no longer clinically
appropriate.
Check the patient and provide any necessary
clinical care.
Change the alarm limit if it is no longer clinically
appropriate.
Check the patient and provide any necessary
clinical care.
Check the ECG lead being used to calculate the
heart rate (the top displayed lead) – make sure
that the QRS amplitude on this lead is at least
0.5 mV.
Change to another ECG lead to get adequate
QRS amplitude.
Reposition or change electrodes if no lead gives
adequate QRS amplitudes. Remember the
importance of good skin preparation techniques.
Check the patient and provide any necessary
clinical care.
Check the ECG lead being used to calculate the
heart rate (the top displayed lead) – make sure
that the QRS amplitude on this lead is at least
0.5 mV.
Change to another ECG lead to get adequate
QRS amplitude.
Reposition or change electrodes if no lead gives
adequate QRS amplitudes. Remember the
importance of good skin preparation techniques.
Check to make sure electrodes are still securely
attached to the patient, and reattach if
necessary. Remember the importance of good
skin preparation techniques.
Check to make sure all the lead wires are still
connected to the electrodes.
Check to make sure the lead wires are securely
connected to the monitor.
Check to make sure there are no broken lead
wires.
Turn monitor off, then back on
If message persists, contact Zoe technical
support.
Press ALARM SILENCE in the event of intentional
removal by clinician.
Calm the patient.
Isolate the patient from auxiliary equipment, if
possible.
Check to make sure electrodes are still securely
attached to the patient, and reattach if
necessary. Remember the importance of good
skin preparation techniques.
740
SELECT
Page 25 of 40