IBP Monitoring
WARNING – Operating the Nightingale PPM3 with P1 and/or P2 alarm
limits disabled means that no low or high P1 and/or P2 alarm
conditions will product alarm notifications. Use this feature with
extreme caution. Patients must be closely observed if P1 and/or P2
limit alarms are disabled.
WARNING – Ensure that no part of the patient connections touches
any electrically conductive material including earth.
WARNING – Only use invasive pressures transducers that can
withstand defibrillation as required by ANSI/AAMI BP22 standard.
WARNING – Mechanical shock to the invasive blood pressure
transducer may cause severe shifts in zero balance and calibration,
and cause erroneous readings.
Note – To reduce nuisance alarms, the invasive pressure alarms are inactivated for a few
seconds during the zeroing process. Invasive pressure alarms are reactivated shortly
after the zeroing process has been completed.
9-4
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e