9.3. IBP Monitoring Messages

Note – Factory default IBP alarm and limit settings may be found in the PPM3 Monitor
Settings chapter.
Note – Substitute P2 for P1 as required in the following table.
Message
P1s < [lower limit]
P1m < [lower limit]
P1d < [lower limit]
P1s > [upper limit]
P1m > [upper limit]
P1d > [upper limit]
P1 signal out of range [low]
P1 signal out of range [high]
P1 unplugged
Unable to zero
Zero required
Zero required and 60 seconds
has expired
Calibration required
Calibration in progress
Cannot calibrate
ART: Check Transducer
April 2019
Parameter
Possible Cause
Value
The patient's invasive
systolic/diastolic/mean
[number]
pressure has fallen
below the current lower
alarm limit.
The patient's invasive
systolic/diastolic/mean
[number]
pressure has risen above
the current upper alarm
limit.
The invasive signal is out
--
of range.
IBP cable or transducer
--
not plugged
--
Stopcock is not open
--
IBP channel not zeroed
--
IBP channel not zeroed
IBP channel requires
--
calibration
IBP channel calibration
--
is in progress
--
Calibration failed
ART IBP pressure below
--
10mmHg
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
IBP Monitoring
Suggested Action
Check the patient and provide any
necessary clinical care.
Verify that the values are not due to
artifact by checking the position of the
patient, cables and transducer. Zero set
the pressure if necessary.
Change the alarm limit if it is no longer
clinically appropriate.
Check the patient and provide any
necessary clinical care.
Verify that the values are not due to
artifact by checking the position of the
patient, cables and transducer. Zero set
the pressure if necessary.
Change the alarm limit if it is no longer
clinically appropriate.
Verify the position of the patient, cables
and transducer. Zero set the pressure if
necessary
Verify that the cable and transducer are
properly connected.
Press ALARM SILENCE if intentionally
unplugged by clinician.
Open the stopcock and check tubing and
cables
Zero the IBP channel
Zero the IBP channel
Arrange for calibration service of the IBP
channel
Wait until the calibration process is
completed
Contact Zoe Medical technical support
Check ART catheter to ensure that it is
properly positioned and connected
9-5