Integrated Pulmonary Index; Checking The Co Signal - ZOE PPM3 User Manual

Nightingale monitoring system
Table of Contents

Advertisement

Oridion CO
should appear in yellow beneath the top channel ECG lead – replacing the
impedance respiration parameter box and waveform area.
2. Select the proper Oridion CapnoLine® or FilterLine® single-use sampling
line based on clinical guidelines and hospital standard of care.
3. Open the cover labeled Microstream® ETCO
4. Place the orange or yellow threaded connector of the sampling line into
the CO
input port beneath the opened cover.
2
a. Gently turn the connector clockwise into the CO
secure connection is achieved. DO NOT OVER TIGHTEN. Over
tightening may warp the connector such that ventilation or the CO
measurement may be compromised. This assures that there is no
leakage of gases during measurement at the connection point and that
measurement accuracy is not compromised.
b. When the connector is secure, the CO
"CO2 Warming Up" message will briefly appear.
5. Securely connect all components and check connections for leaks
according to standard clinical procedures.
6. Connect the patient applied end of the sampling line to the patient as per
instructions included with the CapnoLine ® or FilterLine® packaging.
7. The CO
waveform and numeric parameter values should appear within
2
approximately 10 seconds.
7.3. Checking the CO
When you have connected the patient following the steps listed above, you should
be able to see a CO
waveform on the PPM3 display as shown below. The PPM3
2
should also display values for the patient's end-tidal carbon dioxide (ETCO
fractional inspired carbon dioxide (FICO
Integrated Pulmonary Index (IPI), and alarm limit settings. The infrared light
source used in the Microstream® module generates only the specific wavelengths
characteristic of the CO
required when different concentrations of N
vapor are present in the patient's gases.
7-2
Monitoring
2
Signal
2
), respiration rate from CO
2
absorption spectrum. Therefore, no compensations are
2
O, O
2
N i g h t i n g a l e P P M 3 U s e r ' s G u i d e
on the PPM3 left side.
2
input port until a
2
pump should start and the
2
(RRc),
2
, anesthetic agents, and water
2
2
),
2
April 2019