ECG MONITORING
HR PARAMETER SETUP
Table 11 on page 38 lists the Default HR Alarm Limits for Adult, Pediatric, and Neonatal.
Table 12 on page 39 lists the Default ECG & HR Parameter Setting for Adult, Pediatric, and
Neonatal. HR Alarm Limits and Parameter settings will operate for the current monitor patient
mode.
HR ALARM LIMITS:
1) Touch HR Numeric field and the Setup HR menu will be displayed (refer to Figure 12);
2) Set Alarms On to Yes to enable HR alarms, or No to disable HR alarms.
If Alarms On is set to "No" the monitor will not generate any visual or audible
Warning:
indication of an alarm condition for HR.
3) Adjust the desired HR Upper or Lower Limit value.
The HR Upper and Lower Alarm Limits can be adjusted independently.
4) Touch OK to accept or Cancel to ignore the selection.
5) Touch the Home touch area to return to the Main screen.
HR AUTO (set) ALARM LIMITS
The Upper and Lower alarm limits for HR can be set automatically using the Auto feature,
which provides a structure to establish alarms limits based upon your patient's current
measured value.
To enable Auto limits, open the HR Setup Window and touch the "AUTO" key on the same
line with the parameter (refer to Table 1 for HR percentages).
HR
21-22-0335 Rev D

Figure 12: HR Setup Menu

(x% of the current measured value
Parameter

Table 1: HR Auto Alarm Limit Adjustment

Alarm Limit Adjustment
Lower
80%
740
Upper
125%
Page 21 of 40