Tracking Generator Level Flatness: E4402B, E4403B, E4404B, E4405B,
Tracking Generator Level Flatness: E4402B,
E4403B, E4404B, E4405B, E4407B and E4408B,
Option 1DN
Test Limits
Flatness ≤ 10 MHz
Flatness > 10 MHz
Test Description
A calibrated power sensor is connected to the tracking generator output
to measure the power level at 50 MHz. The power meter is set for REL
mode so that future power level readings are in dB relative to the power
level at 50 MHz. The tracking generator is then stepped to several
frequencies throughout its range. The output power difference relative
to the power level at 50 MHz is measured at each frequency and
recorded.
Required Equipment
Instrument
Meters
Power Meter
RF Power Sensor
Adapters
Type-N (f) to BNC
(m),75-ohm
Chapter 3
E4407B and E4408B, Option 1DN
Minimum
−3.5 dBm
−2.5 dBm
Critical
Specifications
(for this test)
Frequency Range:
100 kHz to 3.0 GHz
Functional Testing
Maximum
+3.5 dBm
−2.5 dBm
Recommended
HP/Agilent Model
438A or E4418A,
E4419A
8482A
1250-1534
3-49