Firmware Errors

Firmware Errors

Table 5-10. Firmware Errors
Error Number
Error Description [description/explanation/examples]
30000
desc = "Internal Communications Error"
help = "An internal software error has occurred involving communications
between different software processes. Please record this error including the
extra text and cycle power. If the error persists, contact the nearest Agilent
Technologies Instrument support center for assistance. In the U.S., call (800)
403-0801. See the Agilent 86140B series Users Guide for a listing of the
Agilent sales and service offices."
30001
desc = "Auto-Measure Software Error"
help = "An error has been detected in the auto-measure software. There is an
internal problem with the software. Please make a note of the text in
parentheses at the end of the error message and cycle power. If the error
persists, contact the nearest Agilent Technologies Instrument support center for
assistance. In the U.S., call (800) 403-0801. See the Agilent 86140B series Users
Guide for a listing of the Agilent sales and service offices."
5-38