Table 7-13. Option 001 Current Source
Current Output
Range
Resolution
a
Accuracy
Clamp Voltage
b
a
Noise Density at 1 kHz
Stability Within 30 Minutes
a
Temperature Drift
Pulse Mode
Pulse Range
Pulse Resolution
Duty Cycle Range
a. Characteristic
b. Nominal
Specifications and Regulatory Information
Specifications
a
7-16
Agilent 86140B, 86141B, 86142B, 86146B Only
±
0 to
200 mA (source or sink)
µ
50
A steps
±
µ
2%
50
A
±
2.7V
<4 nA/
Hz
±
<100 ppm
500 nA
±
°
<(100 ppm
500 nA)/
C
µ
10
s to 6.5 ms
100 ns
Pulse width/1 s to 100%