Agilent Technologies 86142B User Manual page 97

Optical spectrum analyzer
Table of Contents

Advertisement

N o t e
Key Path
Related Functions
Remote
Commands
Key Path
Function Reference
Calibrator Multi-Pt Align
Calibrator Multi-Pt Align
Adjusts the mechanical position of the instrument's internal optical compo-
nents ensuring amplitude accuracy of your measurements. Before initiating
the alignment, connect the internal calibrator to the front-panel input connec-
tor.
Refer to "External Multipoint Wavelength Calibration" on page 2-17
The instrument automatically sets the start wavelength at 1490 nm, stop
wavelength at 1590 nm, span, and reference level, and then performs a fully
automatic, internal auto align. The input signal is aligned at equally spaced
alignments (minimum 50 nm spacing between points) for the internal, multi-
mode fiber or (minimum 5 nm spacing between points) for the external, sin-
gle-mode fiber (Agilent 86144B/86146B only).
Error 5056, Trajectory align cannot find input signal, will occur if a broadband
light source is not connected to the front-panel input connector.
Error 5060, Trajectory align failed, will occur if the align procedure failed.
System > More System Functions > Service Menus > Adv Service Functions
> More Adv Service Menu > Multi-Point Align > Calibrator Multi-Pt Align
Auto Align & Add To Trajectory
Auto Align Preset
User Source Multi-Pt Align
CALibration:ALIGn:INTernal
Center Wavelength (WL)
Sets the center wavelength using the knob, step keys, or numeric keys. The
span remains constant. The center wavelength and the start and stop wave-
length settings are related as follows:
Center Wavelength
Wavelength > Center WL
3-18
stop wavelength
=
-------------------------------------------------------------------------------------------------------- -
+
start wavelength
2