None of the Automeasure Setup parameters are affected by the front-panel
Preset key. They are not saved as part of the measurement setup.
Auto Meas
Auto Align
DISPlay[:WINDow[1]]:TRACe:ALL[:SCALe][:AUTO]
DISPlay[:WINDow[1]]:TRACe:X{:SCALe]:AUTO:SPAN
DISPlay[:WINDow[1]]:TRACe:Y{:SCALe]:AUTO:PDIVision
DISPlay[:WINDow[1]]:TRACe:ALL[:SCALe][:AUTO]:MARKer OFF|ON|0|1
Auto Ranging (Amplitude Setup Panel)
When on, the auto ranging allows the OSA to achieve the desired sensitivity by
taking several sweeps and switching between gain stages. The final trace data
is a blended composite of each trace taken in the different gain stages.
The sweep time (ST) that is displayed in the lower portion of the display is the
time that it takes for the OSA to sweep over one gain stage. The OSA may take
up to three sweeps in three different gain stages to make the measurement.
This depends on the settings for sensitivity, reference level, auto range and also
the particular device being measured. The final data trace is a blended
composite of each trace taken in the different gain stages. For more
information,
refer to "Sweep Time" on page
Setting Auto Range to off will force the OSA to use a single gain stage and
achieve the desired sensitivity by applying a digital filter.
Amplitude > Amplitude Setup > Auto Range
SENSe:POWer:DC:RANGe:AUTO
Function Reference
Auto Ranging (Amplitude Setup Panel)
3-89.