Trace Integ
Trace Integ
Calculates total power. Total power is the summation of the power at each
trace point, normalized by the ratio of the trace point spacing and the resolu-
tion bandwidth. The analyzer can only calculate the total power of single
trace. For example, if a total power calculation is being performed on trace A,
turning total power calculation for trace B will turn the calculation for trace A
off.
Markers > More Marker Functions > Line Marker Menu > Advanced Line
Mkr Functions > Trace Integ
Integrate Limit, Sweep Limit, Search Limit
CALCulate:TPOWer:STATe
Trace Math
Accesses the menu that allows you to manipulate and compare traces by add-
ing or subtracting trace-amplitude and display-line data. Trace Math charac-
terizes changes due to environmental stimulus and displays the cumulative
effect of multiple devices. When using trace math, the traces must be obtained
using identical wavelength, scale and amplitude values. Any difference in ref-
erence level, amplitude units-per-division, or amplitude units invalidates any
data resulting from trace math.
Drift" on page 2-29
The following functions can be accessed:
• All Math Off
• Default Math Trace C
• Default Math Trace F
• Exchange Menu
• Trace A Offset
Traces > Trace Math Off
3-94
See "Using Trace Math to Measure Wavelength