Table 7-4. Dynamic Range Specifications
In 0.1 nm resolution
,
bandwidth
b
c
1250 to 1610 nm (chop mode on)
±
±
±
,
0.5 nm,
1 nm,
5 nm
d
±
1550 nm at
0.8 nm
±
(
100 GHz at 1550 nm)
f
±
1550 nm at
0.5 nm
g
±
(
62.5 GHz at 1550 nm)
±
g
1550 nm at
0.4 nm
±
(
50 GHz at 1550 nm)
±
d
1550 nm at
0.2 nm
±
(
25 GHz at 1550 nm)
a. The 86144B and 86146B specifications are for the 50
b. Excluding multiple order grating response
c. With applied input fiber 9/125
d. Characteristic
e. Chop mode not available on the 86144B, 86146B models
f. Average of all states of polarization
g. Characteristic for 86140B, 86141B, 86143B, 86140B with option 025, 86143B with option 025
Table 7-5. Monochromator Input Specifications
Monochromator Input
,
Input Return Loss
b
c
a. The 86144B and 86146B specifications are for the 50
b. Depends on the quality of the attached connector
µ
c. With a 9/125
m straight connector
Agilent 86140B, 86143B,
a
86144B
–70 dB
e
–60 dB
–58 dB
–55 dB
–40 dB
µ
m internal path only.
µ
m
>35 dB
µ
m internal path only.
Specifications and Regulatory Information
Agilent 86142B, 86145B,
86146B
a
–70 dB
–60 dB
–58 dB
–55 dB
–40 dB
Agilent 8614xB
a
>35 dB
Specifications
Agilent 86141B and
86140B Option 025,
86143B w/Opt 025
–70 dB
–60 dB
–55 dB
–52 dB
----
>35 dB