Agilent Technologies 86142B User Manual page 88

Optical spectrum analyzer
Table of Contents

Advertisement

Key Path
Related Functions
Remote
Commands
Key Path
Related Functions
Remote
Commands
When AMPCOR is turned on, the correction points are applied across the
active measurement range and added to all measurement results. Between
points, the correction values are interpolated linearly or logarithmically. When
measuring at wavelengths outside the first and last correction points, the first
or last value (as appropriate) is used as the correction value.
Whenever AMPCOR is active, the currently selected correction set is dis-
played in the lower left corner of the screen. For example, if correction set
number one is selected, "A1" is displayed.
Remote Commands" on page 5-16
remote commands.
Amplitude > Amplitude Setup...
Reference Level, Sensitivity
(Amplitude Display Mode):DISPlay:WINDow:TRACe:Y:SCALe LINear
DISPlay:WINDow:TRACe:Y:SCALe LOGarithmic
(Amplitude Units):UNIT:POWer
(Auto Zero):CALibration:ZERO
(Reference Level Position):DISPlay:WINDow:TRACe:Y:SCALe:RPOSition
Amplitude Units (Amplitude Setup Panel)
Selects the amplitude scale's units. Selecting Auto for amplitude units allows
power to be shown in dB when using the logarithmic scale and in watts when
using the linear scale. Selecting W forces the readout to be in watts, regardless
of the amplitude scaling.
Amplitude > Amplitude Setup > Amplitude Units
Display Mode, Trace Marker, Scale/Div
UNIT:POWer
Amplitude Units (Amplitude Setup Panel)
Refer to "Amplitude Correction
for an overview of the amplitude correction
Function Reference
3-9