N o t e
Key Path
Remote
Commands
Key Path
Related Functions
Remote
Commands
Recall Menu
Accesses the Recall Setup panel for the following settings:
• Recall (all measurement data or trace data only)
• Recall From (recall from a floppy disk or from the internal drive
• Network File Share 1|2|3|4
• Network File Path
To insure accurate measurements, a wavelength calibration should be
performed each time measurement data is recalled from memory.
Save/Recall > Recall Menu
MMEMory:LOAD:TRACe
Reference Level
Specifies the reference level value, the maximum expected power to be mea-
sured, at the reference level position. The reference level position is indicated
on the display by a dashed green line and the REF annotation on the display.
The default position is one major graticule division from the top of the display.
The reference level is entered in dBm. Use the Amplitude Setup panel to
change the reference level position.
Setup Panel)" on page 3-70
Amplitude > Reference Level
Amplitude Setup (Reference Level Position), Scale/Div
DISPlay:WINDow:TRACe:Y:SCALe:RLEVel
See "Reference Level Position (Amplitude
Function Reference
Recall Menu
3-69